Micro Ohm Meter

Specifications

RANGE 1 Micro Ohms to 19.99K ohm is 8 full scale ranges of 1999 micro ohms, 19.99 milli ohms, 199.9 milli ohms, 1.999 ohms, 19.99 ohms,199.9 ohms, 1.999K ohms and 19.99 K Ohms
Resolution 1 Micro Ohm in 1999 Micro Ohms range. / 0.01ohm in 19.99 ohms range
  0.01 milli ohm in 19.99 milli ohm range. / 0.01ohm in 199.9 ohms range
  0.1 milli ohm in 199.9 milli ohm range. / 0.001ohm in 1.999 ohms range
  1 milli ohm in 1.999 ohms range. / 0.01K ohms in 19.99K ohms range
ACCURACY ± 0.02% ± least counts in all ranges except ±0.5% ±3 least counts in 200 milli ohms, 20 milli ohms and 2000 micro ohms ranges
TEST CURRENT 1.5 Amp DC in 1999 micro ohms / 10 milli Amp DC in 19.99 ohms range.
  1 Amp DC in 19.99 milli ohms, 199.9 milli ohms / 1 milli Amp DC in 199.9 ohms range.
  100 milli Amp DC in 1.999 ohms range. / 100 micro Amp DC in 1.999K ohms range.
  10 micro Amp DC in 19.99 K ohms range.
DISPLAY 3 digit 7 segment LED display with over-range indication. In orver-range conditions, 1 appears in the left display and the remaining 3 displays are turned off.
POWER REQUIREMENT 230 volts AC ± 10%. 50 Hz, 10 VA.
CABINET SIZE & WEIGHT 100 (H) x 220 mm (W) x 210mm (D). 4 Kgs.
STANDARD ACCESSORIES 2 pairs bananna to crocodile test leads and instruction manual.
Need Help?